Jandel RM3-AR Test Unit with PC Software
Combined constant current source and digital voltmeter designed especially for the four point probe measurement. Measure sheet resistance in the range from 1 milliohm-per-square up to 5 x 108 ohms-per-square. Volume resistivity measurement range from 10-3 to 106 ohms-cm. System accuracy within 0.3%. The RM3 Test Unit is now the RM3-AR Test Unit with input current auto-ranging! The RM3-AR will calculate the optimum input current range automatically.
|
 (click image to see larger photo)
 Shown here displaying result in ohms-per-square
|
SPECIFICATIONS:
Superior Current Source
- 10nA to 100mA (99.999mA) current source selectable in steps to 5 decimal place resolution
- Current set numeric keypad
- 4 default preset current programs (user programmable)
Superior Inbuilt DVM
- Input Impedance 1,000,000,000,000 ohms
- Input Bias current 4pA
- Two Ranges: Hi - 0 to 150mV and
Low - 0 to 1250mV
- 150mV Range Accuracy: 0.2% +/- 5uV
resolution (10uV or 1uV) range
- 1250mV Range Accuracy: 0.2%+/-100uV
resolution (100uV Ohms/Square)
- Rapid Zeroing null function for DVM
Download a one page product brochure (83K pdf file) regarding the RM3-AR Test Unit
Download the instruction manual (687k pdf file) for the RM3-AR Test Unit
- software is included with RM3-AR - see pages 10 through 16 in the instruction manual for details. The RM3-AR will read-out directly in ohms-per-square without using the software. The software allows measurement data to be converted to ohms-cm, and allows saving the measurement data in a CSV format file.
|
Features
- 28 Key high quality Keypad
- 16x2 line LCD Display for simultaneous indication of Set Current and either Ohms/Sq or mV
- Intuitive operation
- Microprocessor controlled
- Reduced Footprint
- Robust Attractive ABS Case
- Accurately measures down to 10's of milliohms/square without external meter
- 4mm socket facility to connect an external meter
- RS232/USB connectivity for control and for collecting data in CSV format
- Automatic compliance voltage limit protects your sample which reduces maximum voltage as higher currents are set 5-40V
- Warning messages when compliance limit is reached, or DVM input exceeded
- Self test facility
- Flash upgradeable software
- CE Marked
- AD/DA converters and amplifiers for the DVM and setting the Current source and DVM, are Burr Brown 110V or 240V operation (selectable)
|
A complete four point probing system consists of the RM3-AR Test Unit shown above, combined with one of the mechanical four point probe units:
Multiposition Wafer Probe
Multi Height Probe
Microposition Probe
Multi Height Microposition Probe
Universal Probe
Hand Applied Probe
|